Description
CA 8:1
ENDO
Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.
Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.
Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.
CA 8:1 Endo | |
---|---|
Transmission ratio | 8:1 |
Maximum speed (rpm) | 5’000 |
Head diameter (mm) | 8.5 |
Head size (mm) | 9.4 |
Weight (g) | 46 |
Coupling system | ISO 3694 and Bien-Air Micro-Series |
Warranty | 1 to 2 years 1 |
Reference | CA 8:1 Endo 1600911-001 |
PREPARATION
A.REMOVING THE INSTRUMENT
Remove the rotary instrument from the chuck mechanism
Disconnect the contra-angle from the motor
CLEANING
B.REMOVING DIRT/DEPOSITS
Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
Shake the Spraynet before use
Use the Spraynet attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)
Do not immerse the instrument in a disinfectant bath
OR
LUBRICATION
A.LUBRICATION WITH LUBRIFLUID
Shake the Lubrifluid before usen
Use the Lubrifluid attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out