PMP 10:1

Handpieces with interchangeable heads and contra-angles with detachable cups are the only instruments that can be fitted to motors rotating at 40,000 rpm. Apart from superior comfort and extended service life,Bien-Air prophylaxis instruments in particular offer high resistance to brasive cleaning pastes and can withstand constant and repeated sterilizations.

Description

PMP 10:1 PROPHY

Handpieces with interchangeable heads and contra-angles with detachable cups are the only instruments that can be fitted to motors rotating at 40,000 rpm. Apart from superior comfort and extended service life,Bien-Air prophylaxis instruments in particular offer high resistance to brasive cleaning pastes and can withstand constant and repeated sterilizations.

Specifications

TECHNICAL DATA

PMP 10:1 PROPHY
Type Straight handpiece for standard disposable heads.
Ratio 10:1 (speed reducing factor of 10)
Cleaning / sterilisation Thermal washer / steam autoclave up to 134°C (273°F)
Warranty 2 to 3 years 1
Reference PMP 10:1 PROPHY 1600289-001

Maintenance

MAINTENANCE OF CONTRA-ANGLES

PREPARATION

A.REMOVING THE INSTRUMENT

Remove the rotary instrument from the chuck mechanism

Disconnect the contra-angle from the motor

B.PRE-DISINFECTION

External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

A.CLEAN UNDER RUNNING WATER AT <38°C

B.REMOVING DIRT/DEPOSITS

Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire

Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

Shake the Spraynet before use

Use the Spraynet attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Clean the exterior of the contra-angle with a cloth soaked in Spraynet

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

A.LUBRICATION WITH LUBRIFLUID

Shake the Lubrifluid before usen

Use the Lubrifluid attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant

Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

STERILISATION

A.
PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated