With its pre-programmed operating sequences and streamlined ergonomics, the Chiropro L is the perfect partner tor your implantology and dental surgery procedures. The Chiropro L features 8 of the main implant systems on the market. At each stage of the procedure, the screen displays all of the clinical data you need to ensure your operations go smoothly. Thanks to the irrigation system built into the instrument, the Chiropro L facilitates handling and prevents heating of the bone.

Description

CHIROPRO L UNBEATABLE PERFORMANCE.

With its pre-programmed operating sequences and streamlined ergonomics, the Chiropro L is the perfect partner tor your implantology and dental surgery procedures. The Chiropro L features 8 of the main implant systems on the market. At each stage of the procedure, the screen displays all of the clinical data you need to ensure your operations go smoothly. Thanks to the irrigation system built into the instrument, the Chiropro L facilitates handling and prevents heating of the bone.

The power and precision of the Chiropro L from Bien-Air enable you to confidently overcome any challenge!

Technology

MXi_LED

MX-i LED Micromotor

High torque for unrivalled working comfort, at both high and low speeds. With ceramic ball bearings to guarantee exceptional service life and reliability.

CA-20-1

CA 20:1 L Micro-Series
Contra-Angle

Fits into the physiodispenser’shandpiece’s internal irrigation system for smooth and efficient performance and improved service life. Equipped with a remarkably small head and a double LED system for uniform lighting regardless the speed of rotation.

PM-1-2

PM 1:2 Straight Handpiece

Designed to withstand the stresses caused from repeated lateral pressure. With a new bur guard to prevent the physiological liquid from returning into the physiodispenser – for impeccable hygiene.

foot pedal

Multifunctional Foot Pedal

To command the micromotor’s speed and rotation variation, toggle the peristaltic pump, and control the operation sequences.

irrigation lines

Disposable Irrigation Lines

Bien-Air has designed a patented peristaltic pump. Our disposable irrigation lines incorporate an insertion support ensuring simplified handling and perfect hygiene.

Specifications

TECHNICAL DATA

Chiropro L iChiropro Chiropro 3rd Gen
Control Knob (sterilisable)
Implant placement
Built-in surgical procedures
Surgical protocols 8 >300
Guided surgery protocols
Digital workflow (import data from planning software CodiagnostiXTM)
Exportable surgical report
Max. Torque at the rotary instrument in implant mode 55 Ncm 70 Ncm 70 Ncm
Dimensions L x W x H 309mm x 220mm x 123mm 242mm x 244mm x 102mm 240mm x 240mm x 102mm
Weight 2,7 Kg 2,8 Kg 2,2 Kg
Device warranty 2 years 2 to 3 years 1 2 to 3 years 1
MX-i specifications MX-i LED 1600755-001
Self-ventilated
MX-i LED 1600755-001
Self-ventilated
MX-i LED 1601008-001
Non self-ventilated
Micromotor Dimensions ø x L 23 x 91mm 23 x 91mm 23 x 83mm
Micromotor Weight 115g 115g 110g
Reference Chiropro L set with CA 20:1 L
1700307-001Chiropro L set without CA
1700298-001
iChiropro set with Micro-Series
CA 20:1 L
1700440-001iChiropro set without Micro-Series
CA 20:1 L 1700439-001
Chiropro 3rd Gen set without CA
1700708-001Chiropro 3rd Gen set with
Micro-Series CA 20:1 L
1700707-001

Maintenance

MAINTENANCE OF CONTRA-ANGLES AND HANDPIECES FOR SURGERY AND IMPLANTOLOGY

PREPARATION

A.REMOVING THE INSTRUMENT

Remove the rotary instrument from the chuck mechanism

Disconnect the contra-angle from the motor

B.PRE-DISINFECTION

External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

A.CLEAN UNDER RUNNING WATER AT <38°C

B.REMOVING DIRT/DEPOSITS

Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire

Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.INTERNAL CLEANING WITH AQUACARE

Shake the Aquacare before use

Use the Aquacare attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds

Use the Aquacare attachment 3

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second

D.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

Shake the Spraynet before use

Use the Spraynet attachment

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Clean the exterior of the contra-angle with a cloth soaked in Spraynet

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

A.LUBRICATION WITH LUBRIFLUID

Shake the Lubrifluid before use

Use the Lubrifluid attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant

Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

STERILISATION

A.PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated