Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.

Description

CA 8:1
ENDO

Bien-Air CA 8:1 Endo contra-angle is designed to work with shorter endo file shanks. Offering unparalleled visibility in molar areas, its small head provides the ideal solution for rotary endodontic procedures.

Specifications

TECHNICAL DATA

CA 8:1 Endo
Transmission ratio 8:1
Maximum speed (rpm) 5’000
Head diameter (mm) 8.5
Head size (mm) 9.4
Weight (g) 46
Coupling system ISO 3694 and Bien-Air Micro-Series
Warranty 1 to 2 years 1
Reference CA 8:1 Endo 1600911-001

Maintenance

MAINTENANCE OF CONTRA-ANGLES

PREPARATION

A.REMOVING THE INSTRUMENT

Remove the rotary instrument from the chuck mechanism

Disconnect the contra-angle from the motor

B.PRE-DISINFECTION

External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

A.CLEAN UNDER RUNNING WATER AT <38°C

B.REMOVING DIRT/DEPOSITS

Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire

Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

Shake the Spraynet before use

Use the Spraynet attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Clean the exterior of the contra-angle with a cloth soaked in Spraynet

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

A.LUBRICATION WITH LUBRIFLUID

Shake the Lubrifluid before usen

Use the Lubrifluid attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant

Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

STERILISATION

A.PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated