Description

NEW CHIROPRO PURE SIMPLICITY

Simplicity. The cornerstone of the new range of implant and oral surgery motors developed by Bien-Air Dental. General dental practitioner or experienced surgeon, are you seeking a fast, simple and efficient solution for your implant dentistry procedures? Look no further than the Chiropro. Your perfect solution for precise and smooth placement of implants.

Technology

THE 7 ESSENTIAL CHARACTERISTICS OF NEW CHIROPRO.

Specifications

TECHNICAL DATA

Chiropro 3rd Gen Chiropro L iChiropro
Control Knob (sterilisable)
Implant placement
Built-in surgical procedures
Surgical protocols 8 >300
Guided surgery protocols
Digital workflow (import data from planning software CodiagnostiXTM)
Exportable surgical report
Max. Torque at the rotary instrument in implant mode 70 Ncm 55 Ncm 70 Ncm
Dimensions L x W x H 240mm x 240mm x 102mm 309mm x 220mm x 123mm 242mm x 244mm x 102mm
Weight 2,2 Kg 2,7 Kg 2,8 Kg
Device warranty 2 to 3 years 1 2 years 2 to 3 years 1
MX-i specifications MX-i LED 1601008-001
Non self-ventilated
MX-i LED 1600755-001
Self-ventilated
MX-i LED 1600755-001
Self-ventilated
Micromotor Dimensions ø x L 23 x 83mm 23 x 91mm 23 x 91mm
Micromotor Weight 110g 115g 115g
Reference Chiropro 3rd Gen set without CA
1700708-001Chiropro 3rd Gen set with
Micro-Series CA 20:1 L
1700707-001
Chiropro L set with CA 20:1 L
1700307-001Chiropro L set without CA
1700298-001
iChiropro set with Micro-Series
CA 20:1 L
1700440-001iChiropro set without Micro-Series
CA 20:1 L 1700439-001

Video

Maintenance - MX iMOTOR

CLEANING

A.INTERNAL CLEANING

Rinse the interior of the piece under running water at &tl;38°C

B.EXTERNAL CLEANING

Clean the exterior of the piece under running water at <38°

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT
(follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION
(follow the manufacturer’s instructions for use)

STERILISATION

A.PLACE THE MOTOR IN A STERILISATION BAG

B.

TERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Maintenance - Contra Angles

PREPARATION

A.REMOVING THE INSTRUMENT

Remove the rotary instrument from the chuck mechanism

Disconnect the contra-angle from the motor

B.PRE-DISINFECTION

External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

A.CLEAN UNDER RUNNING WATER AT <38°C

B.REMOVING DIRT/DEPOSITS

Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire

Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.INTERNAL CLEANING WITH AQUACARE

Shake the Aquacare before use

Use the Aquacare attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds

Use the Aquacare attachment 3

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second

D.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

Shake the Spraynet before use

Use the Spraynet attachment

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Clean the exterior of the contra-angle with a cloth soaked in Spraynet

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

A.LUBRICATION WITH LUBRIFLUID

Shake the Lubrifluid before use

Use the Lubrifluid attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant

Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

STERILISATION

A.PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated