NEW CHIROPRO PLUS

Simplicity. The cornerstone of the new implant motor developed by Bien-Air Dental. General dental practitioner or experienced surgeon, are you seeking a fast, simple and efficient solution for your implant dentistry procedures? Look no further than the Chiropro. Your perfect solution for precise and smooth placement of implants.

Description

NEW CHIROPRO PLUS PURE SIMPLICITY

Simplicity. The cornerstone of the new implant motor developed by Bien-Air Dental. General dental practitioner or experienced surgeon, are you seeking a fast, simple and efficient solution for your implant dentistry procedures? Look no further than the Chiropro. Your perfect solution for precise and smooth placement of implants.

Technology

THE 9 ESSENTIAL CHARACTERISTICS OF NEW CHIROPRO PLUS.

Specifications

TECHNICAL DATA

Chiropro PLUS 3rdGen Chiropro 3rd Gen iChiropro
Control Knob (sterilisable)
Implant placement
Built-in surgical procedures
Surgical protocols >300
Guided surgery protocols
Digital workflow (import data from planning software CodiagnostiXTM)
Exportable surgical report
Max. Torque at the rotary instrument in implant mode 80 Ncm 70 Ncm 70 Ncm
Dimensions L x W x H 240mm x 240mm x 102mm 240mm x 240mm x 102mm 242mm x 244mm x 102mm
Weight 2,2 Kg 2,2 Kg 2,8 Kg
Device warranty 2 to 3 years 1 2 to 3 years 1 2 to 3 years 1
MX-i specifications MX-i LED 1600755-001
Self-ventilated
MX-i LED 1601008-001
Non self-ventilated
MX-i LED 1600755-001
Self-ventilated
Micromotor Dimensions ø x L 23 x 91mm 23 x 83mm 23 x 91mm
Micromotor Weight 115g 110g 115g
Reference Chiropro PLUS 3rd Gen set
without CA 1700710-001Chiropro PLUS 3rd Gen set with
Micro-Series CA 20:1 L
1700709-001Chiropro PLUS 3rd Gen set with
Micro-Series CA 1:2.5 L
1700751-001
Chiropro 3rd Gen set without CA
1700708-001Chiropro 3rd Gen set with
Micro-Series CA 20:1 L
1700707-001
iChiropro set with Micro-Series
CA 20:1 L
1700440-001iChiropro set without Micro-Series
CA 20:1 L 1700439-001

Video

Maintenance - Contra Angles

CONTRA ANGLES AND HANDPIECES (SURGERY AND IMPLANTOLOGY)

PREPARATION

A.REMOVING THE INSTRUMENT

Remove the rotary instrument from the chuck mechanism

Disconnect the contra-angle from the motor

B.PRE-DISINFECTION

External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

A.CLEAN UNDER RUNNING WATER AT <38°C

B.REMOVING DIRT/DEPOSITS

Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire

Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

C.INTERNAL CLEANING WITH AQUACARE

Shake the Aquacare before use

Use the Aquacare attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds

Use the Aquacare attachment 3

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second

D.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

Shake the Spraynet before use

Use the Spraynet attachment

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second

Clean the exterior of the contra-angle with a cloth soaked in Spraynet

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)

LUBRICATION

A.LUBRICATION WITH LUBRIFLUID

Shake the Lubrifluid before use

Use the Lubrifluid attachment 1

Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second

Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant

Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

STERILISATION

A.PLACE THE CONTRA-ANGLE IN A STERILISATION BAG

B.

STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONSInstruments bearing the “135°C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated

 

Maintenance - MX iMOTOR

CLEANING

A.INTERNAL CLEANING

Rinse the interior of the piece under running water at &tl;38°C

B.EXTERNAL CLEANING

Clean the exterior of the piece under running water at <38°

DISINFECTION

A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

B.THERMAL DISINFECTION
(follow the manufacturer’s instructions for use)

STERILISATION

A.PLACE THE MOTOR IN A STERILISATION BAG

B.

TERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

Instruments bearing the “135°C” symbol can be sterilised in an autoclave

Downloads

MANUAL

MAINTENANCE

IMAGES

MX IMOTOR

Reviews

There are no reviews yet.

Be the first to review “NEW CHIROPRO PLUS”