Simplicity. The cornerstone of the new implant motor developed by Bien-Air Dental. General dental practitioner or experienced surgeon, are you seeking a fast, simple and efficient solution for your implant dentistry procedures? Look no further than the Chiropro. Your perfect solution for precise and smooth placement of implants.
Technology
THE 9 ESSENTIAL CHARACTERISTICS OF NEW CHIROPRO PLUS.
Maintenance - Contra Angles
CONTRA ANGLES AND HANDPIECES (SURGERY AND IMPLANTOLOGY)
PREPARATION
A.REMOVING THE INSTRUMENT
Remove the rotary instrument from the chuck mechanism
Disconnect the contra-angle from the motor
B.PRE-DISINFECTION
External disinfection
Do not immerse the instrument in a disinfectant bath.
CLEANING
A.CLEAN UNDER RUNNING WATER AT <38°C
B.REMOVING DIRT/DEPOSITS
Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire
Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.INTERNAL CLEANING WITH AQUACARE
Shake the Aquacare before use
Use the Aquacare attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds
Use the Aquacare attachment 3
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second
D.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
Shake the Spraynet before use
Use the Spraynet attachment
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
Clean the exterior of the contra-angle with a cloth soaked in Spraynet
DISINFECTION
A.DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer’s instructions for use)
Do not immerse the instrument in a disinfectant bath
OR
B.THERMAL DISINFECTION (follow the manufacturer’s instructions for use)
LUBRICATION
A.LUBRICATION WITH LUBRIFLUID
Shake the Lubrifluid before use
Use the Lubrifluid attachment 1
Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant
Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
STERILISATION
A.PLACE THE CONTRA-ANGLE IN A STERILISATION BAG
B.
STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONSInstruments bearing the “135°C” symbol can be sterilised in an autoclave
Only sterilise instruments which have been cleaned and lubricated